extend your senses
KOS
TEMPERATURE SENSORS FOR OUTDOOR USE WITH A PLASTIC CONNECTION HEAD

TEMPERATURE SENSORS FOR OUTDOOR USE WITH A PLASTIC CONNECTION HEAD

S 110 series

TEMPERATURE SENSORS FOR OUTDOOR USE WITH A PLASTIC CONNECTION HEAD

CONFIGURATOR

CONFIGURE AND REQUEST

20765
__KOUPIT__
__NABIDKA__
__JINY DOTAZ__

BBB

--%>

These resistance-type temperature sensors are intended for contact temperature measurement of gaseous substances. The plastic connection head is provided with a cable outlet bushing (terminal board is located in the head) or a connector. The standard temperature range in which the sensors are allowed to be utilised is -50 to 100 °C. The temperature sensors can be utilised for any control systems that are compatible with sensing element output signals or output signals quoted in the table of sensing element types. Easy mounting of the temperature sensor is ensured by the unique “S head” design invented by SENSIT s.r.o.

The temperature sensors are designed to be operated in a chemically non-aggressive environment.

ACCESSORIES

DECLARATION, CERTIFICATION, CALIBRATION

  • Manufacturer provides EU Declaration of Conformity.
  • Calibration – The final metrological inspection – comparison with standards or working instruments – is carried out for all the products. Continuity of the standards and working measuring instruments is ensured within the meaning of the Section 5 of Act no.505/1990 on metrology. The manufacturer offers a possibility to supply the sensors calibrated in SENSIT s.r.o.’s laboratory (according to requirements of the EN ISO/IEC 17025 standard) or in an Accredited laboratory.  

Caution: the temperature sensors with the output 4–20 mA can be delivered with the new connection head only in the version A.

Technické parametry


Sensor type
(K - with connector)
NS 110x
NS 110xK
NS 111x
NS 111xK
NS 112x
NS 112xK
NS 310x
NS 310xK
NS 311x
NS 311xK
Type of sensing element
Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range
-50 to 100°C
Maximum measuring
DC current
1 mA 1 mA 1 mA 0.3 mA 0.3 mA


Sensor type
(K - with connector)
NS 113x
NS 113xK
PTS 110x
PTS 110xK
PTS 210x
PTS 210xK
PTS 310x
PTS 310xK
HS 110x
HS 110xK
Type of sensing element T1 = Ni 226 PT 100/3850 PT 500/3850 PT 1000/3850 termistor NTC 20 kΩ
Measuring range -50 to 100 °C
Maximum measuring
DC current
0.7 mA 3 mA 1.5 mA 1 mA 0.3 mW *)

*) maximum power consumption

Sensor type
(K - with connector)
NS 510A
NS 510AK
NS 710x
NS 710xK
Note
Type of sensing element Pt 1000/3850 PT1000/3850
Output signal
4 to 20 mA 0 to 10 V
Measuring renges
-30 to 60 °C
0 to 35 °C
0 to 100 °C
0 to 150 °C
-30 to 60 °C
0 to 35 °C
0 to 100 °C
0 to 150 °C
Ambient temperature arround the connection -30 to 70 °C
Power supply (U) 11 to 30 V DC 15 to 30 V DC Recommended value 24 V DC
Load resistance Rz 150 Ω for power supply 12 VDC
700 Ω for power supply 24 VDC
> 10kΩ
Output signal
- sensing elements braek
> 24 mA > 10.5 mA
Output signal
- sensing element short
< 3.5 mA ~ 0 V

Note: x = version A or version B

**)According to the customer‘s requirement, it is possible to provide a customized measuring range from -40 to 150 °C; the minimum span of the range must be 35 °C (e.g. -20 to 15 °C; -30 to 80 °C)

OTHER PARAMETERS
Accuracy class
Ni sensing elements: B class, t = ± (0.4 + 0.007t), for t ≥ 0; t = ± (0.4 + 0.028|t|), for t ≤ 0 in °C;
Pt sensing elements: B class according EN 60751, t = ± (0.3 + 0.005|t|) in °C
NTC 20 kΩ: ± 1 °C for the range 0 to 70 °C
Measuring error
NS 510A(K), NS 710x(K)
NS 510A ± 1 °C (based on air flow speed)
NS 710A 0,6 % for range with span; min. 0,5 °C
Sensors connection
according to the wiring diagram
Standard length of the stem (version A) for resistance output and for output 0 to 10 V : 25 mm
for output 4 to 20 mA : 50 mm
Time response
τ0.5 < 9 s (in air flow at 1m s-1) – version A
τ0.5 ≤ 30 s (in air flow at 1m s-1) – version B
Recommended wire cross section
– sensors with grommet
0.35 to 1.5 mm2
Type of connector in the head
– sensor with connector
RSFM4 - Lumberg
Insulation resistance
> 200 MΩ at 500 V DC, 25° ± 3 °C; humidity < 85 %
Ingress protection IP 65 in accordance with EN 60529
Material of the stem
stainless steel DIN 1.4301 – version A
Material of the connection head
POLYAMIDE
Operating conditions
ambient temperature: -50 to 100 °C; -30 to 70 °C with a converter
relative humidity: max. 100 % (at the ambient temperature 25 °C)
atmospheric pressure: 70 to 107 kPa
Weight approximately 0.15 kg

Note: Manufacturer reserves the right to make changes in design and technical characteristics of the products. For actual information, please always contact your dealer.




Rozměrový náčrt

Schéma zapojení

Modifikace

  • option of encasing two sensing elements  
  • option of encasing non-standard tempeature sensors (DALLAS, TSic, KTY, SMT, etc.)  
  • accuracy class A (with the exception of sensors Ni 10000/5000, Ni 10000/6180, T1 = Ni 2226, termistor NTC 20 kΩ)
  • option of three- or four- wire connection
  • possibility of providing custom temperature ranges for temperature sensors with converter

Soubory ke stažení

008.18en.pdf 008.18en.pdf

Related Products

TEMPERATURE SENSORS FOR OUTDOOR USE WITH DISPLAY

Detail Detail

TEMPERATURE SENSORS FOR OUTDOOR USE WITH A METAL CONNECTION HEAD

Detail Detail

TEMPERATURE AND HUMIDITY SENSORS FOR INTERIOR AND OUTDOR USAGE WITH 4 to 20mA OUTPUT

Detail Detail
1
 
2
 
3
*
 
4
*

CONTACT

SENSIT s.r.o.

Školní 2610, 756 61 Rožnov pod Radhoštěm


+420 571 625 571

b6a3eacd-1e06-4d71-ac5e-b164dbb27053

NEWSLETTER

Receive news and product information in your email.

© Copyright 2022 SENSIT s.r.o. | Školní 2610, 756 61 Rožnov pod Radhoštěm, Czech Republic | +420 571 625 571 | 32b660dd-e8eb-45c2-9fa9-f88b4a31601a
All Rights Reserved. Webdesign: FlexiSystems s.r.o.: e-learning, webdesign, CMS.
...